64 DF-760E
User Interface
to trip if the pressure exceeds preset limits. The user may assign the pressure alarm to one
of four relays.
Figure 44: Moisture Pressure Range Alarm Setup
The limits are not user adjustable but are set automatically on the basis of the background
gases entered in the GSF screen. See page 103 for additional information on setting the
background gases.
Use the
Next
key to move from between fields and use the arrow keys (
and
to
change the highlighted selections and to enter numerical values. When done, use the
Next
key to move to the
Accept
button and hit the
Enter
key to return to the main display.
Using the
ESC
key at anytime will exit the screen making no changes and return to the
main display.
7.5.6.4 Moisture System Alarm Setup
Figure 45: Moisture System Alarm Setup
If left for more than 30 minutes, a System Error Code will trip a System Alarm if
configured to do so. See page 60 for additional information on System Error Codes.
NOTE: The 30 minute clock is delayed for 60 minutes on a “cold” system start up.
Summary of Contents for NanoTrace DF-760E
Page 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Page 4: ......
Page 38: ......
Page 44: ......
Page 122: ......
Page 128: ......
Page 138: ......
Page 142: ......
Page 143: ...Service DF 760E 139...
Page 144: ...140 DF 760E...
Page 148: ......
Page 166: ......
Page 168: ......
Page 171: ...DF 760E 167...
Page 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Page 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...