74 DF-760E
User Interface
7.6.2 Oxygen Calibrate Menu
Figure 63: Oxygen Calibrate Menu
7.6.2.1 Check/Adjust Zero
Figure 64: Check/Adjust Zero Screen
It is important to note that sample line contributions to oxygen offsets in ppb
measurements, especially at the sub-ppb level, are difficult to remove even with a correctly
operating system. The dry down process of the analyzer and/or gas supply system can
literally take many months. The oxygen analyzer has an array of zeroing features that
enable the user to establish
temporary
performance near 0.0 ppb during the dry down
process. Monitoring the zero reference number during that time, or repeated restoration of
the factory zero, will give the user a sense of the offset being induced by the user zero
actions.
This Check/Adjust Zero screen Figure 64 enables the user to pick from numerous zero
calibration related functions.
The
Zero Reference
value in the upper right hand corner is a reflection of the deviation
of the oxygen sensor’s present zero current from the original factory calibration based on
any user zero actions. Instruments are shipped from the factory with the zero reference set
to 0.00. After a manual or auto zero is performed, the zero reference may change.
Summary of Contents for NanoTrace DF-760E
Page 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Page 4: ......
Page 38: ......
Page 44: ......
Page 122: ......
Page 128: ......
Page 138: ......
Page 142: ......
Page 143: ...Service DF 760E 139...
Page 144: ...140 DF 760E...
Page 148: ......
Page 166: ......
Page 168: ......
Page 171: ...DF 760E 167...
Page 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Page 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...