106 DF-760E
User Interface
7.7.4 Gas Delivery
Figure 117: Gas Delivery
The Gas Delivery screen allows for monitoring of the gas delivery system valves. An ‘X’
indicates that the valve is closed. This screen is for display purposes only and allows no
user interaction.
Figure 118: Gas Delivery System Status
7.7.4.1 Back Flow Prevention System
The Back Flow Prevention System provides a means of eliminating the possibility of
contamination of the plumbing system when gas flow is lost or accidentally shut off.
Under this condition ambient air may be drawn in to the system through the bypass exhaust
by means of the aspirator or vacuum pump.