SENS.413347.015RE
97
K.87 Description of input parameters of calibration functions (182), their designations
and access restriction are shown in table K.6
Table K.6
Name
Access
Description of parameter of calibration command
P
A
B
1
-
+
+ Reset of alarm locking signaling
2
-
+
+
Adjustment of lower measurement threshold of primary
variable
3
-
+
+
Adjustment of upper measurement threshold of primary
variable
4
-
+
+ Initialization of primary transducer
5
-
+
+ Restore parameters LL and rS
6
-
+
+ Enable mode of gas detector indication check
7
-
+
+ Disable mode of gas detector indication check
8
-
+
+ Calibration of output current 4 mA of 4...20 mA interface
9
-
+
+ Calibration of output current 20 mA of 4...20 mA interface
10
-
+
+ Save value of current during calibration of interface 4...20 mA
11
-
+
+
Initialization of factory calibration settings for primary
transducer
12
-
+
+
Set lower measurement threshold (lower value of range,
corresponding to 4 mA, analogue of command 37)
13
-
+
+
Set upper measurement threshold (upper value of range,
corresponding to 4 mA, analogue of command 36)
230
+
+
+ Enter user level
231
+
+
+ Enter administrator level