SENS.413347.015RE
80
Cont. of table K.1
#
Function
Message format
15
Read device
information
Query data
No
Response data
Byte
Type
Value
Description
0
enum
-
Code of alarm signal type:
0
– alarm current is more than 12
mA;
1
– alarm current is less or equal to
12 mA.
1
enum
0
Code of transmission function
(linear)
2
enum
57
Code of range measurement units
(%)
3-6
float
-
Upper level of range (corresponds
with 20 mA)
7-10
float
-
Lower level or range (corresponds
with 4 mA)
11-14
float
-
Damping value (s)
15
enum
-
Write-protection code:
0x00
– unlocked;
0x01
– locked temporarily (until
power reset);
0x02
– permanently locked;
16
enum
Manufacturer's code
17
bits
1
Flag of analogue channel
connection (always connected)
Error codes
Code
Description
16
Access restricted
16
Read final
assembly
number
Query data
No
Response data
Byte
Type
Value
Description
0-2
u-24
-
Final assembly number
Error codes
Code
Description
16
Access restricted