SENS.413347.015RE
23
N
o.
Item
Subitem (parameter)
Note
Designati
on
Name
Desig
nation
Name, unit of measurement
С09
Calibration of output current
20 mA of 4...20 mA interface
С10
Save value of current during
calibration of interface 4...20
mA
Cont. of table 8.3
No
.
Item
Subitem (parameter)
Note
Designati
on
Name
Desig
nation
Name, unit of measurement
5
CAL.
Calibration
C11
Initialization of factory
calibration settings for
primary transducer
Set in accordance with
8.5.3.6.
С200
Exit emulation mode
Set in accordance with
8.4.2.16.
С201
Enter emulation mode
С210
Exit verification mode
Performed in accordance
with 8.5.3.5, 8.5.4.
C211
Enter verification mode
Performed in accordance
with 8.5.3.5, 8.5.4
C222
Factory reset
Performed in accordance
with 8.4.2.17
С230
Exit administrator
mode
Performed in accordance
with 8.4.2.6.
After changing gas
detector settings in
operation mode with
administrator access level,
to
save
changes
, it is
necessary to run command
C230
.
С231
Enter administrator
mode
6
PSd.
Access control
P1
Password to enter mode for
changing parameters
(administrator mode)
Set in accordance with
8.4.2.6.
7
SADA
Configuration of
RS-485 port
parameters
rS
Data rate
Set in accordance with
8.4.2.14.
rP
Parity
SP
Protocol switch