SENS.413347.015RE
26
gas detector models with methane calibration gas which has factory setting of parameters
LL
and
rS
corresponding to
methane
detected component (LL = 4.4; rS = 1)).
Table 8.4
Detected component
Value of parameter LL (LFL of component, %, vol. fraction)
Propane (С
3
H
8
)
1.7
Butane (С
4
Н
10
)
1.4
Hexane (С
6
Н
14
)
1.0
Note - LFL of component in accordance with GOST R IEC 60079-20-1-2011.
After selection of detected component, initialization of primary transducer must be
performed, in accordance with 8.4.2.8.
As a result of primary transducer initialization, values of parameters
LL
and
rS
will be
matched to serial number of PT and saved into nonvolatile memory of gas detector.
Note - matching of value of parameters LL and
rS
to serial number of PT is used as
protection function against unauthorized changing of parameters
LL
,
rS
and replacement of
PT. In case of changing of value of parameters
LL
and
rS
or replacement of PT without further
initialization, gas detector will indicate failure in accordance with table 8.5...
In case of damage of value of parameters
LL
and
rS
, for example, due to operator's
mistake, they can be restored by running command
C05
. Recovery of parameters
LL
and
rS
is
only possible if initialization of primary transducer was not performed with changed
parameters.
Value of parameters
LL
and
rS
must be recorded in the table of configuration parameters
of gas detector configuration data sheet.
8.4.2.8 Initialization of primary transducer and viewing of its serial number.
To initialize PT, "administrator" access level is required.
Initialization of PT is performed as follows:
- enter configuration menu;
- select
CAL
menu item;
- run command
C231
(enter administrator mode);
- run command
C04
(initialization of primary transducer);
- run command
C230
(exit administrator mode).
As a result of initialization, parameters
s1
and
s2
of
Cont
settings menu item will be
assigned following values:
-
s1
- highest four digits of PT serial number;
-
s2
- lowest four digits of PT serial number.
Viewing of parameters
s1
and
s2
is performed as follows:
- enter configuration menu;
- select
Cont
menu item;
- select and view parameter
s1
;
- select and view parameter
s2
.
Viewing parameters
s1
and
s2
is available with "user" and "administrator" access levels.
Value of parameters
s1
,
s2
must be recorded in the table of configuration parameters of
gas detector configuration data sheet.
8.4.2.9 Configuration of alarm signaling thresholds.
Alarm signaling thresholds are configured in menu item
LEu.
On the basis of configured
thresholds, status byte of gas detector is formed (see Annex H). When specified threshold