SENS.413347.015RE
42
Hysteresis
. Viewing of "Hysteresis" parameter value is available for operation modes
with "user" and "administrator" access level. Setting of "Hysteresis" parameter value is
performed in operation mode with "administrator" access level.
1. Enter operation mode with "administrator" access level.
2. Double-click with LMB on line of "Hysteresis" parameter.
3. In the appeared dialogue box (figure 8.15) enter required value in "Value" input field
and click "OK" button.
Figure 8.15 - "Setting of parameter" dialogue box. Input of "Hesteresis" parameter value
4. Write parameter by pressing "F4" keyboard key.
If all done correctly, message "Parameters successfully written" will appear in event log
and flag "Configuration was changed" will be set. Factory setting of "Hysteresis" parameter is
5.000%.
Signal
relays
. Two parameters are used for configuration of signal relays: "Relay 1
settings" (b1) and "Relay 2 settings" (b2). Viewing of parameters values is available for
operation modes with "user" and "administrator" access level. Setting of events for relay
triggering is performed in operation mode with "administrator" access level.
1. Enter operation mode with "administrator" access level.
2. Double-click with LMB on "Relay 1 settings" or "Relay 2 settings" parameter line.
3. Dialogue box "Setting of parameter" appears on the screen (Figure 8.16).
Figure 8.16 - "Setting of parameter" dialogue box. Input of "Relay 1 settings" parameter value
4. To receive values for signal relays settings is as follows (values of bits of signal
relays are shown in Annex I).
For example, you need to active signal relay 1 for triggering upon exceeding of 50% LFL
value of threshold 5. To do this, set bit 5 (figure 8.17).
1
2
3
4
5
6
7
8
0
0
0
0
1
0
0
0