Ion concentration
ProLab 4000
116
ba75646e03
12/2012
12
In the
Standard
register, enter the values for
Standard volume
and
Standard concentration
if necessary.
13
Use
[Dosing]
to automatically add to the sample the adjusted
volume of the standard solution (
Standard volume
)
or
If the burette control is inactive:
Add to the sample the adjusted volume of the standard solution
manually (
Standard volume
)
14
Start the measurement with
[Continue]
.
The measured value is checked for stability (Stability control).
15
Wait for the measurement with stability control to end.
or
Take over the calibration value with
[OK]
.
The
Calculation
register with the calculated measured value
pops up.