ProLab 4000
Conductivity
151
ba75646d03
12/2012
Note
The measuring method can only be started if
Ref. temp.
,
Start temp.
and
Temp. interval
are selected so that the reference temperature cor-
responds to the nominal temperature of a measuring point:
Ref. temp.
=
Start temp.
+ x*
Temp. interval
6
Warm or cool the test sample to the start temperature
Start
temp.
.
As soon as the temperature of the measuring point is achieved
and a stable value is recognized, the measurement is automat-
ically started and the dataset stored.
After the measurement, a piece of information about the next
measuring point with the relevant nominal temperature is dis-
played and the temperature difference between the current
temperature and the nominal temperature.
7
Slowly warm or cool the test sample until the end temperature
End temp.
is achieved.
Each time the nominal temperature for a measuring point is
reached a dataset is stored.
The corresponding calibration record with the determined tem-
perature coefficients is displayed at the end of the calibration.
If the calibration is successful, the determined temperature
dependency for the
TC nLin2
method is stored and used for fur-
ther measurements with this method.
8
Confirm the record with
[Continue]
.
[nLin2] is displayed in the status line of the measuring channel.
Summary of Contents for ProLab 4000
Page 4: ...ProLab 4000 4 ba75646e03 12 2012...
Page 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Page 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Page 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Page 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Page 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Page 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Page 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Page 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Page 227: ......