17.2
Measuring ranges, resolution, accuracy
17.2.1 Temperature
17.2.2 pH/ORP
Measuring ranges,
resolution
Variable
Measuring range
Resolution
T [°C]
- 35.0 ... + 150.0
0.1
Accuracy
(± 1 digit)
Variable
Accuracy
Temperature of the
test sample
T [°C]
± 0.1
- 35.0 ... + 150.0
Manual
temperature input
Variable
Range
Increment
T
manual
[°C]
-35 ... + 150
1
Measuring ranges,
resolution
Variable
Measuring range
Resolution
pH
- 2.000 ... + 20.000
- 2.00 ... + 20.00
- 2.0 ... + 20.0
0.001
0.01
0.1
U [mV]
- 2200.0 ... + 2200.0
- 2200 ... + 2200
0.1
1
Accuracy
(± 1 digit)
Variable
Accuracy
Temperature of the test
sample
pH
± 0.002
± 0.01
± 0.1
+ 15 °C ... + 35 °C
U [mV]
/ range
- 2200.0 ... + 2200.0
± 0.1
+ 15 °C ... + 35 °C
- 2200 ... + 2200
± 1
+ 15 °C ... + 35 °C