pH
ProLab 4000
72
ba75646e03
12/2012
Criteria
With identical measurement conditions, the following applies:
Note
You can terminate prematurely the
Stability control
function with
<OK>
manually at any time. If the
Stability control
function is termi-
nated prematurely, the current measurement data is not downloaded
to the interface.
5
Using
<SC>
or
<MODE>
, release the frozen measured value.
The [SC] status display disappears.
Measured param-
eter
Reproducibility
Response time
pH value
Better than 0.01
> 15 seconds