ORP voltage
ProLab 4000
90
ba75646e03
12/2012
8
.2
Measuring the ORP
8
.2.1
Preparatory activities
Perform the following preparatory activities when you want to measure:
Temperature sensor
You can measure with or without a temperature sensor. The
TP
display
indicator appears when a temperature sensor is connected.
1
Adjust the temperature of the solutions and measure the cur-
rent temperature if the measurement is made without a temper-
ature sensor.
2
Connect the ORP electrode to the meter.
3
Highlight the channel of the electrode in the measured value
display.
4
If necessary, select the U or dU display with
<MODE>
.
5
Adjust the temperature of the solutions and measure the cur-
rent temperature if the measurement is made without a temper-
ature sensor.
6
With manual temperature input:
Call up the manual input of the temperature value with
Measuring
/
Setting of man. temp...
.
7
Enter the measured temperature value and confirm it with
[OK]
.
8
Check the meter with the electrode.
Note
ORP electrodes are not calibrated. However, you can check ORP
electrodes using a measuring solution.
Note
If using an ORP electrode without temperature sensor, you can also
use the temperature sensor of another electrode (see page 62). In this
case, the indication from which channel the temperature value is
taken appears next to the temperature value.
Summary of Contents for ProLab 4000
Page 4: ...ProLab 4000 4 ba75646e03 12 2012...
Page 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Page 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Page 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Page 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Page 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Page 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Page 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Page 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Page 227: ......