Ion concentration
ProLab 4000
130
ba75646e03
12/2012
9.6.6
Sample subtraction
In the
Sample subtraction
procedure, a known amount of test sample
is added to the standard solution.
The ion concentration in the sample is calculated from the change in
potential.
The sample subtraction is one method for the indirect determination of
ions. It is used for the determination of ions for which no direct determi-
nation is possible.
Calibration
Measuring
1
Check the currently displayed ion.
If the ion to be measured is displayed incorrectly:
Select the ion type in the menu,
Measuring
/
ISE methods
/
Ion
type
.
2
In the
Measuring
/
ISE methods
menu, select the method,
Sample subtraction
.
3
Carry out a two-point calibration according to the user guid-
ance (see page 106).
As soon as a stable value is achieved for the second standard
calibration solution, the slope (mV) and evaluation (Error, +++)
of the electrode after calibration is displayed.
The infotext,
Accept calibration and terminate
is displayed.
4
Using
[OK]
, confirm the infotext
Accept calibration and termi-
nate
.
The calibration record is displayed and output to the interface.
The calibration procedure is finished.
5
Close the calibration record with
[OK]
.
The
ISE method
: The
Sample subtraction
window pops up.
The
Standard
register is displayed.
Summary of Contents for ProLab 4000
Page 4: ...ProLab 4000 4 ba75646e03 12 2012...
Page 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Page 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Page 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Page 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Page 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Page 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Page 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Page 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Page 227: ......