
ProLab 4000
Ion concentration
119
ba75646e03
12/2012
5
Thoroughly rinse the electrode with deionized water.
6
Immerse the electrode in the sample.
7
Start the measurement with
[Continue]
.
After end of the measurement, the
Standard
pops up.
8
In the
Standard
register, enter the value for
Standard concen-
tration
if necessary.
9
Use
[Dosing]
to automatically add to the sample the adjusted
volume of the standard solution (
Standard volume
) (1 vol% of
sample solution).
or
If the burette control is inactive:
Add to the sample the adjusted volume of the standard solution
manually (
Standard volume
) (1 vol% of sample solution).
Summary of Contents for ProLab 4000
Page 4: ...ProLab 4000 4 ba75646e03 12 2012...
Page 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Page 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Page 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Page 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Page 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Page 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Page 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Page 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Page 227: ......