ORP voltage
ProLab 4000
94
ba75646e03
12/2012
As an alternative you can also determine the difference of the ORPs of
two solutions with one ORP electrode only. To do so, first define the
ORP of one solution as the zero point.
Stability control
The
Stability control
function checks the stability of the measurement
signal. The stability has a considerable impact on the reproducibility of
measured values. The background color is red until the measured
value is stable.
Regardless of the setting for
Autom. stability control
(see page 63) in
the
System
menu, you can start the
Stability control
function manually
at any time.
1
Perform the preparatory activities (see page 69).
2
Prepare the reference solution for the determination of the ref-
erence point.
3
Immerse the electrode in the reference solution.
4
Highlight the channel of the electrode in the measured value
display.
5
If necessary, select the U display with
<MODE>
.
6
Display the voltage of the current zero point with
<CAL>
or
Measuring
/
Referenz
/
Determine reference...
.
7
Measure the reference solution with
<OK>
.
The measured voltage is defined as the zero point.
or
Terminate the display of the zero point with
<ESC>
.
8
Rinse the electrode and immerse it into the test sample.
Summary of Contents for ProLab 4000
Page 4: ...ProLab 4000 4 ba75646e03 12 2012...
Page 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Page 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Page 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Page 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Page 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Page 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Page 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Page 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Page 227: ......