What to do if...
ProLab 4000
200
ba75646e03
12/2012
Warning
[TempErr]
Warning
[ISEErr]
Burette does not dose
16.3
Conductivity measurement
Conductivity measuring
cell was not recognized
Error message,
Error
(invalid calibration)
Cause
Remedy
– Temperature difference
between measurement and cali-
bration greater than 2 K.
– Adjust the temperature of
the test sample
Cause
Remedy
– Electrode voltage outside cali-
brated range
– Recalibrate
Cause
Remedy
– Address of the burette incor-
rectly set
– Enter the address of the
burette on the meter
(see page 101)
– Enter the address of the
burette on the burette
(see operating manual of the
burette)
– RS232 interface of the burette
incorrectly set
– Configure the RS232 inter-
face of the burette
(see page 29 and operating
manual of the burette)
Cause
Remedy
– Measuring cell not connected
– Connect measuring cell
– Cable broken
– Replace measuring cell
Cause
Remedy
– Measuring cell contaminated
– Clean cell and replace it if
necessary
– Unsuitable control standard
– Use control standard,
0.01 mol/l KCl
Summary of Contents for ProLab 4000
Page 4: ...ProLab 4000 4 ba75646e03 12 2012...
Page 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Page 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Page 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Page 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Page 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Page 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Page 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Page 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Page 227: ......