Transmitting data
ProLab 4000
1
8
6
ba75646e03
12/2012
13.2.2 USB-B interface
(
USB Device
)
Connect the interface to the PC via the supplied Z875 USB cable.
Installation of the USB
driver on the PC
System requirements of the PC for installation of the USB driver:
PC with Pentium processor or higher with at least one free USB con-
nection and CD-ROM drive
Windows 2000, XP.
Note
If the burette control is active, data output to the RS232 interface
(printer) is blocked.
Note
Pin assignment of the socket: See page 203.
1
Insert the supplied installation CD in the CD drive of your PC.
2
Connect the meter to the USB-B input (
USB Device
) of the PC
via the USB cable.
Windows automatically installs the driver for the meter. Follow
the Windows installation instructions as necessary.
The meter is listed as a virtual COM interface among the con-
nections in the Windows instrument manager.
3
Set up similar transmission data on the PC/printer:
Baud rate: selectable in the range 1200 ... 115200
Handshake: RTS/CTS + Xon/Xoff
Set on the PC only:
– Parity: none
– Data bits: 8
– Stop bits: 1
Summary of Contents for ProLab 4000
Page 4: ...ProLab 4000 4 ba75646e03 12 2012...
Page 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Page 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Page 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Page 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Page 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Page 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Page 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Page 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Page 227: ......