ProLab 4000
What to do if...
201
ba75646e03
12/2012
CalClock
with red frame
Cond temp?
16.4
General errors
Meter does not react to
keystroke
Time is lost
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring
system
Cause
Remedy
– Test sample does not have the
correct temperature (max.
±
2
°C temperature difference from
calibration temperature)
– Adjust the temperature of the
test sample
Cause
Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the
<On/Off>
and
<SC>
key simultaneously.
Cause
Remedy
– Buffer batteries empty
– Exchange buffer batteries