pH
ProLab 4000
70
ba75646e03
12/2012
7.2
Measuring the pH value
7.2.1
Preparatory activities
Perform the following preparatory activities when you want to measure:
Temperature sensor
You can measure with or without a temperature sensor. The
TP
display
indicator appears when a temperature sensor is connected.
1
Adjust the temperature of the solutions and measure the cur-
rent temperature if the measurement is made without a temper-
ature sensor.
2
Connect a pH electrode to the meter.
3
Highlight the channel of the pH electrode in the measured
value display.
4
If necessary, select the pH display with
<MODE>
.
5
With manual temperature input:
– Call up the input window for the manual temperature input
with
Measuring
/
Setting of man. temp...
.
– Measure the temperature of the test sample using a ther-
mometer.
– Enter the measured temperature value and confirm it with
[OK]
.
6
Calibrate the meter with the electrode.
Note
An incorrect calibration of pH electrodes leads to incorrect measured
values. Calibrate regularly. The meter has to be calibrated before
measuring.
Note
If using a pH electrode without temperature sensor, you can also use
the temperature sensor of another connected electrode or conductiv-
ity measuring cell (see page 62). In this case, the indication from
which channel the temperature value is taken appears next to the tem-
perature value.
Summary of Contents for ProLab 4000
Page 4: ...ProLab 4000 4 ba75646e03 12 2012...
Page 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Page 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Page 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Page 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Page 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Page 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Page 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Page 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Page 227: ......