ProLab 4000
pH
77
ba75646e03
12/2012
Stability control
The calibration procedure automatically activates the stability control
function. The current measurement with stability control can be termi-
nated at any time (accepting the current value).
Calibration record
When finishing a calibration, the new calibration values are first dis-
played as an informative message. Then you can decide whether you
want to take over these values of the new calibration or whether you
want to continue measuring with the old calibration data. After accept-
ing the new calibration values the calibration record is displayed.
Display of calibration
data and download to
interface
You can display the data of the last calibration (see page 166). Subse-
quently, you can transmit the displayed calibration data to the interface,
e. g. to a printer or PC, with the
<PRINT>
key.
Sample record:
Note
You can display the slope in the unit,
mV/pH
or
%
(see page 73).
Note
The calibration record is automatically transmitted to the interface
after calibrating.
Print date26.04.07 16 was selected:13
ProLab 4000 (06249876)
Current user: 1234567890
CALIBRATION pH
Calibration date20.04.07 10:14:03
A 161 1M-BNC-ID (A062498765)
User: Administrator
Calibration interval 7 days
AutoCal DIN 2-point TAuto SC
Buffer 1
6.86
Buffer 2
9.18
Voltage 1
7.2 mV 26.3 °C
Voltage 2
-171 mV 26.3 °C
Slope
-59.2 mV/pH
Zero point
-0.5 mV
Sensor
+++
Summary of Contents for ProLab 4000
Page 4: ...ProLab 4000 4 ba75646e03 12 2012...
Page 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Page 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Page 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Page 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Page 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Page 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Page 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Page 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Page 227: ......