RTC
®
5 PC Interface Board
Rev. 1.9 e
7 Basic Functions for Scan Head and Laser Control
132
innovators for industry
Notes
• The LASER1 and LASER2 signals are activated by
setting the pulse length to a nonzero value and
are deactivated by setting the pulse length to
zero (default setting).
• The signals LASER1 and LASER2 have a constant
relative phase shift of 180° (half an output
period). LASER2 can be used for the control of a
second laser tube. With
(bit#1=1), both signals can be exchanged with
each other. To control laser power, the pulse
length of the LASER1 and LASER2 signals can be
varied. Both signals share the same pulse lengths
and output frequencies.
• Note that
half
of the output period must be
specified for the LASER1 and LASER2 signals, i.e.
the shift between the two laser signals.
• The commands
list
commands. They can be
used within a list, for instance, for changing the
laser power at any time between two list
commands.
• The actual output period and the pulse lengths of
laser signals LASER1 and LASER2 (but not the
laser delays) are dependent on the time base.
For RTC
®
5 mode, this is fixed at 64 MHz (1 bit
equals 1/64
µ
s).
For RTC
®
4 compatibility mode (for “laser active”
operation), the time base can be set via
(but not via
) for either 1 MHz (1 bit
equals 1
µ
s) or 8 MHz (1 bit equals 1/8
µ
s). For
RTC
®
4 compatibility mode, SCANLAB generally
recommends setting the time base to 8 MHz (via
set_laser_timing(..,..,..,1)
). A time base of
1 MHz should only be chosen if necessary.
The time base for standby signals is always
1/64
µ
s (in RTC
®
5 mode) or 1/8
µ
s (in RTC
®
4
compatibility mode).
The command
(bit#0) can be used
to specify whether a (for “laser active” operation)
started modulation pulse (LASER1 or LASER2) should
execute to completion or be cut off when it has not
yet fully executed when the LASERON signal is
switched off (see
). This cannot be specified
for standby signals; if applicable, these will be cut off.
7.4.4 YAG Modes
( [1, 2, 3 or 5] ) one can choose
between four different YAG laser control modes. The
laser control timing diagram
shows the corresponding signals.
In all four YAG modes, for “laser active” operation (if
applicable after a LaserOn delay and if enabled via
or
)
• The LASERON signal is switched on,
• A
Q-Switch signal
with variable pulse length and
frequency is provided at the LASER1 output and
• A programmable
FirstPulseKiller signal
is
provided at the LASER2 output.
For “laser standby” operation (if applicable after a
LaserOff delay)
• The LASERON signal is switched off and
• Standby pulses are provided at the LASER1
output (but not at the LASER2 output) with pulse
lengths and periods that can be defined via
or
.
The LASER1 signal is activated by setting the pulse
length to a nonzero value and is deactivated by
setting the pulse length to zero (default setting).
Q-Switch Signal
The Q-Switch signal is available for control of the
laser’s Q switch. The Q-Switch period and pulse
length are set with the commands
,
or
.
Notes
• Note that
half
of the Q-Switch output period
must be specified.
• The actual Q-Switch output period and the pulse
length (but not the laser delays) are dependent
on the time base.
For RTC
®
5 mode, this is fixed at 64 MHz (1 bit
equals 1/64
µ
s).
For RTC
®
4 compatibility mode, the time base can
be set via
) for
either 1 MHz (1 bit equals 1
µ
s) or 8 MHz (1 bit
equals 1/8
µ
s). For RTC
®
4 compatibility mode,
SCANLAB generally recommends setting the time
base to 8 MHz (via
set_laser_timing(..,..,..,1)
). A time base of
1 MHz should only be chosen if necessary.