Manual
Weld Seam Detection System R4000 SND40
for Tubes, Cans and Drums
B0059001 / Rev. 2.3
Maintenance
ROLAND ELECTRONIC GmbH · Otto-Maurer-Str. 17 · DE 75210 Keltern · Phone +49 (0)7236-9392-0 · Fax +49 (0)7236-9392-33
141
Start application SND40_10
3: Scroll until you find “SND40_10”
and restart the application.
Method 2: Calibration without PC connection, but with a external USB keyboard
• Connect a USB keyboard.
• Terminate SND40 program (ESC key).
• The procedure in calibration without PC is almost identical.
No Telnet connection is required. So proceed with the calibration as described
in method 1.
• Instead of a USB keyboard you may also use a USB mouse.
• Finally, when you return to the start screen make the taskbar visible.
• Press "Run" and execute "cmd.exe".
• Enter: ndcucfg
• Type the following commands and confirm each by pressing „Enter“::
reg save
quit
The changes are now saved.