Board Replacement
R&S ZVB
1145.1078.82 3.30
E-1
Replacing a Fan
(See Chapter 5, Spare Parts List, Item 15 and drawings 1145.1010, 1145.1290, 1145.1332)
The fans, three in all, are located behind the right-hand side panel.
Opening the instrument and removing the fan
¾
Turn off the instrument and disconnect from the mains, screw off the 4 rear-panel feet (460) and
pull off the enclosure (400) backwards.
¾
Lift off the top instrument cover (296) after undoing the 23 (2 ports) or 28 (4 ports) countersunk
screws (298).
¾
Remove the 4 screws in the front handles (410), right and left, and take off the front handles.
¾
Remove the countersunk screw (390) next to the display and pull off the front cover (300, 303, 306,
310, 313, 316, 320, 323) forwards
¾
Remove the 2 countersunk screws (610) in the top of the front frame and the 2 in the bottom.
¾
Remove the countersunk screws (176) (4 ), (177) (6 ), (178) (8)
¾
Pull out the front unit together with the keyboard and display (600, 601, 602, 603) forwards.
¾
Disconnect the fan cable on the motherboard X35, X36 and X37 (FAN).
¾
Undo the 2 screws (19) and the 5 screws (17) and remove fan plate (50) with the 3 fans (15).
¾
Remove fan (15) by undoing the 4 fan screws (18).
Fitting a new fan and reassembling the instrument
¾
Install the fan using the 4 fan screws.
N.B.:
The arrows on the fan show the installation position. The fan blows air into the instrument.
Route the fan cable so that it cannot get caught in the fan.
¾
Reinsert the fan plate with the 3 fans on it into the instrument and secure with 2 screws (19) and 5
screws (17).
¾
Connect the fan cabling on the motherboard X35, X36 and X37 (FAN).
¾
Reinsert the front unit into the instrument and secure to the front frame with 4 countersunk screws
(610).
¾
Refit the countersunk screws (176) (4), (177) (6 ), (178) (8).
Caution
Avoid trapping cables and ensure cabling is tidy.
¾
Replace the front cover (300 to 323) and secure with the countersunk screw (390).
¾
Refit the 2 front handles (410) using 4 screws.
¾
Refit the top instrument cover (296) with 23 (2 ports) or 28 (4 ports) countersunk screws (298).
¾
Connect the mains cable, turn on at the mains switch and press the ON key.
¾
Check that all three fans are operating correctly (fans are blowing air into the instrument).
¾
Turn off the instrument and disconnect the mains cabling again.
¾
Push the enclosure (400) back on and screw the 4 rear-panel feet (460) into position.
¾
Connect the mains cable, turn on at the mains switch and press the ON key.
Summary of Contents for ZVB 4
Page 57: ...Instrument Construction and Function Description R S ZVB 1145 1078 82 I 3 2 E 1 ...
Page 133: ...1145 1078 82 5 5 E 1 Spare Parts List Mechanical Drawings ...
Page 142: ...1145 1078 82 5 15 E 1 Block Circuit Diagram ...
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 148: ......
Page 149: ......
Page 150: ......
Page 151: ......
Page 152: ......
Page 153: ...1145 1078 82 5 17 E 1 Circuit diagrams ...
Page 154: ......
Page 155: ......
Page 156: ......