R&S ZVB
Manual Alignment and Recording Correction Values
1145.1078.82 2.7
E-1
etc. for the second measurement input.
¾
Set Service Function (Writing to the EProm)
3.1.2.2
When correction value programming
for the two DC voltage measurement inputs has been completed,
end the NWA application and restart.
Check the alignment by applying the four vo1 V, -1 V, +10 V and -10 V and, as a further check,
0 V.
Reading the previous DC values:
¾
Select
Read
in the Service Function Menu
¾
Set Service Functions:
3.1.2.5.dc_meas_1V.DcMeasMultiplier
3.1.2.5.dc_meas_1V.DcMeasOffset
3.1.2.5.dc_meas_10V.DcMeasMultiplier
3.1.2.5.dc_meas_10V.DcMeasOffset