Performance Test
R&S
ZVB
1145.1078.82 1.14
E-1
Checking the DC Measurement Inputs
Test equipment:
DC Power Supply NGSM 32/10
(see Chapter “Test Equipment“, Item9)
Multimeter URE3
(see Chapter “Test Equipment“, Item10)
DC cable
(see Chapter “Test Equipment“, Item11)
Test setup:
Connect the Power Supply to the Input DC MEAS 1V (DC MEAS 10V) of
the ZVB using the DC cable
R&S
®
ZVB settings:
- [
Preset
]
For DC Meas 1 V:
- [
Meas
: More : DC Inputs : DC Meas
±
1 V]
- [
Format
: Real]
- [
Scale
: Scale/Div : .25 x1]
- [
Marker
]
For DC Meas 10 V:
- [
Meas
: More : DC Inputs : DC Meas
±
10 V]
- [
Format
: Real]
- [
Scale
: Scale/Div : 2.5 x1]
- [
Marker
]
Measurement:
Set Power Supply to
DC values U
DC
using Multimeter URE (DC values
see Performance Test Report) and connect it to pos. and neg. input
Read off the DC level U
DC ZVB
indicated by the marker
.
Calculation:
Deviation = U
DC
-
U
DC ZVB