R&S
ZVB Performance
Test
1145.1078.82 1.5
E-1
Checking the Maximum Output Power
Instrument: Power
sensor
or
Power meter with power sensor
(see Chapter “Test Equipment“, Item2)
Test setup:
Connect power sensor to port1 (port 2 to port 4)
Power sensor settings:
For measurement frequencies, see Performance Test Report {f
GEN
*}.
R&S
®
ZVB settings:
- [
Preset
]
- [
Meas
: Wave Quantities: a1 Src Port 1
(a2 Src Port 2 to a4 Src Port 4)]
- [
Sweep
: Sweep Type : CW Mode :
CW Frequency :
{f
GEN
*}
;
Power :
16 dBm
]
- [
Sweep
: Single :
Restart
]
* For measurement frequencies see the Performance Test Report {f
GEN
}.
Measurement:
Read off the level indicated by the power meter.
Max. power ................................................... see Performance Test Report