![Rohde & Schwarz ZVB 4 Manual Download Page 21](http://html.mh-extra.com/html/rohde-and-schwarz/zvb-4/zvb-4_manual_1477900021.webp)
R&S
ZVB Performance
Test
1145.1078.82 1.7
E-1
Checking the Output Linearity
Instrument:
Spectrum analyzer (see Chapter “Test Equipment“, Item1)
Test cable (see Chapter “Test Equipment“, Item7)
Test setup:
Connect the spectrum analyzer to port 1(port 2 to port 4)
Spectrum analyzer settings:
Note:
Synchronize the reference oscillators in the spectrum analyzer and in the
R&S
®
ZVB :
- [
PRESET
]
- [
FREQ
: f
GEN
*]
- [
SPAN
: ZERO SPAN ]
- [
BW
: MANUAL RES BW :
100 Hz
]
- [
MARKER
: DETECTOR : RMS ]
- [
AMPT
: REF LEVEL :
5 dBm
]
* For measurement frequencies, see Performance Test Report {fGEN}.
R&S
®
ZVB settings:
- [
Preset
]
- [
Meas
: Wave Quantities: a1 Src Port 1
(a2 Src Port 2, a3 Src Port 3, a4 Src Port 4)]
- [
Sweep
: Sweep Type : CW Mode :
CW Frequency :
{f
GEN
*}
;
Power :
{l
GEN
*}
]
- [
Sweep
: Single :
Restart
]
* For measurement frequencies {fGEN} and levels {l
GEN
}, see
Performance Test Report
Measurement:
Read the spectrum analyzer’s marker values (level L) .
Calculation:
The measured values are referred to the level at the ZVB setting of
-10 dBm.
Calculating the generator level linearity:
Level linearity = L
- L
@-10dBm
– step width (in
dB)
Step width = l
gen
– (-10 dBm) (reference)
Summary of Contents for ZVB 4
Page 57: ...Instrument Construction and Function Description R S ZVB 1145 1078 82 I 3 2 E 1 ...
Page 133: ...1145 1078 82 5 5 E 1 Spare Parts List Mechanical Drawings ...
Page 142: ...1145 1078 82 5 15 E 1 Block Circuit Diagram ...
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 148: ......
Page 149: ......
Page 150: ......
Page 151: ......
Page 152: ......
Page 153: ...1145 1078 82 5 17 E 1 Circuit diagrams ...
Page 154: ......
Page 155: ......
Page 156: ......