Troubleshooting R&S
ZVB
1145.1078.82 3.52
E-1
Checking the Temperature Sensors
Select Service Level 2 (see Service Functions).
When the
Hardware Info
softkey is pressed
information about the installed hardware, the
results of temperature measurements and …
are displayed.
Temperature Info
Current Temperature Readings
Component
Sensor
Temperature
Sensor
Temperature
Sensor
Temperature
Motherboard
Near NC:
33.75°C Near SY:
34.75°C Near PS:
38.50°C
Netcon 1
Analog:
40.25°C Digital:
49.75°C
Netcon 2
Analog:
----- Digital:
-----
Reflectometer 1
Generator:
38.50°C Receiver:
42.50°C
Reflectometer 2
Generator:
36.00°C Receiver:
38.75°C
Reflectometer 3
Generator:
38.25°C Receiver:
40.25°C
Reflectometer 4
Generator:
39.25°C Receiver:
41.75°C
Highly elevated temperature values indicate a fan failure.