R&S ZVB
Troubleshooting
1145.1078.82 3.59
E-1
Testing the frequency reference board
(see
Test Equipment
)
¾
Remove the board from the instrument.
¾
Reinsert board and extension card.
¾
Connect the output to be tested to the spectrum analyzer using an adapter cable and adapter SMA-
N.
¾
Set the frequencies listed in the table.
¾
Check signals according to following table.
Signal
Connector
Frequency
Nom. level
Setting
NA1_AD_CLK
X103
80 MHz
10 dBm ±1dB
NA2_AD_CLK
X104
80 MHz
10 dBm ±1dB
SY1_REF
X105
128 MHz
10 dBm ±1dB
SY2_REF
X106
128 MHz
10 dBm ±1dB
REF_10_OUTIN
X107
10 MHz
6 dBm ±1dB
Menu/System/Internal
Reference
The board must be replaced if the signal is more than 2 dB below the stated nominal level.
¾
Connect the frequency reference output of the spectrum analyzer to X107 (REF_10_OUTIN) using
an adapter cable and adapter SMA-N.
¾
Connect X105 or X106 (SY1_REF or. SY2_REF) to the spectrum analyzer using an adapter cable
and adapter SMA-N.
¾
The 128 MHz signal’s frequency must precisely equal its nominal frequency.
If the frequency differs in any way (e.g. 128.001 MHz), replace the board.
Summary of Contents for ZVB 4
Page 57: ...Instrument Construction and Function Description R S ZVB 1145 1078 82 I 3 2 E 1 ...
Page 133: ...1145 1078 82 5 5 E 1 Spare Parts List Mechanical Drawings ...
Page 142: ...1145 1078 82 5 15 E 1 Block Circuit Diagram ...
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 148: ......
Page 149: ......
Page 150: ......
Page 151: ......
Page 152: ......
Page 153: ...1145 1078 82 5 17 E 1 Circuit diagrams ...
Page 154: ......
Page 155: ......
Page 156: ......