Performance Test
R&S
ZVB
1145.1078.82 1.10
E-1
Checking the Input Noise Level
Test equipment
Calibration kit
(see Chapter “Test Equipment“, Item4)
Test setup:
Connect the Match Male from the calibration kit to port 1 (port 2 to port 4)
R&S
®
ZVB settings:
- [
Preset
]
- [
Meas
: Wave Quantities: b1 Src Port 1
(b2 Src Port 2, b3 Src Port 3, b4 Src Port 4)]
- [
Power
: RF Off ]
- [
Meas Bandwidth
: 10 Hz ]
- [
Marker ]
- [
Sweep
: Sweep Type : CW Mode : CW Frequency :
{f
GEN
*}
]
- Service Function 1.0.0.1.1 (see chapter 3 ‘Service Functions’, Service
Level 2)
- [
Measure
: Wave Quantities : More Wave Quantities… : Properties :
Detector :
RMS
]
- [
Sweep
: Single :
Restart
]
* For measurement frequencies, see Performance Test Report {fGEN}.
Measurement:
Read off the noise level indicated by the markers on the DUT.