
R&S
ZVB Performance
Test
1145.1078.82 1.13
E-1
Checking the Dynamic Range reduced due to Spurious
Test equipment:
Calibration kit N
(see Chapter “Test Equipment“, Item4)
Test setup:
Connect Short Male to port1 and port 2 (port 3 and port 4)
(use Short Female with Through Male as a second Short Male)
R&S
®
ZVB settings:
- [
Preset
]
- [
Meas
: Ratios : b1/a2 Drive Port 2]
[
Meas
: Ratios: b2/a1 Drive Port 1]
[
Meas
: Ratios: b3/a4 Drive Port 4]
[
Meas
: Ratios: b4/a3 Drive Port 3]
- [
Start
: 16 MHz ]
- [
Number of Points
: 500 ]
- [
Power
: 10 dBm ]
- [
Meas Bandwidth
: 1000 Hz ]
- [
Marker
]
- [
Sweep
: Sweep Type : CW Mode : CW Frequency :
{f
GEN
*}
]
- Service Function 1.0.0.1.1 (see chapter 3 ‘Service Functions’, Service
Level 2)
- [
Measure
: Wave Quantities : More Wave Quantities… : Properties :
Detector :
RMS
]
- [
Sweep
: Single :
Restart
]
* For measurement frequencies, see Performance Test Report {fGEN}.
Measurement:
Read off marker value
Calculation:
Nominal dynamic range: ............................... see Performance Test Report
Summary of Contents for ZVB 4
Page 57: ...Instrument Construction and Function Description R S ZVB 1145 1078 82 I 3 2 E 1 ...
Page 133: ...1145 1078 82 5 5 E 1 Spare Parts List Mechanical Drawings ...
Page 142: ...1145 1078 82 5 15 E 1 Block Circuit Diagram ...
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 148: ......
Page 149: ......
Page 150: ......
Page 151: ......
Page 152: ......
Page 153: ...1145 1078 82 5 17 E 1 Circuit diagrams ...
Page 154: ......
Page 155: ......
Page 156: ......