![Rohde & Schwarz ZVB 4 Manual Download Page 15](http://html.mh-extra.com/html/rohde-and-schwarz/zvb-4/zvb-4_manual_1477900015.webp)
R&S
ZVB Preliminary
Remarks
1145.1078.82 1.1
E-1
1 Performance Test
Preliminary Remarks
•
The required characteristics of the spectrum analyzer are checked after a warm-up time of at least
60 minutes; this ensures that the guaranteed data is met.
•
The values stated in the following sections are not guaranteed data; only the specifications in the data
sheet are binding.
•
The values in the data sheet are guaranteed limits. Because of the measurement errors that arise,
these limits must be increased to encompass the tolerances of the measuring equipment used for
the performance test.
•
Entries for the measurement are represented in the following way:
[
<KEY>
]
Press a front-panel key, e.g. [
SPAN
]
[<SOFTKEY>]
Press a softkey, e.g. [MARKER -> PEAK]
[<
nn unit
>]
Enter a value + terminate the entry with a unit, e.g. [
12 kHz
]
Consecutive entries are separated with a [ : ], e.g. [
BW
: MANUAL RES BW :
3 kHz
]