R&S ZVB
Troubleshooting
1145.1078.82 3.67
E-1
Testing the Motherboard
28 V supply
Using a multimeter, measure the voltage at X 100.B5 (wrt ground).
Permissible deviation: ± 0.5 V
Preamplifier for DC measurement inputs
Apply the DC voltages listed in the table using the 4-pin Mini-DIN connector at the DC measurement
input.
Measure the DC voltage with a multimeter.
Input
Voltage
at
Gnd
APPLIED
VOLTAGE
Measurement at
Rated value
DC MEAS -1 V to +1 V
8
3, 5, 6
- 1 V
X 141.B10
2.33 V
DC MEAS -1 V to +1 V
6
3, 5, 8
+ 1 V
X 141.B11
2.33 V
DC MEAS -10 V to +10 V
8
3, 5, 6
- 10 V
X 141.D10
2.33 V
DC MEAS -10 V to +10 V
6
3, 5, 8
+ 10 V
X 141.D11
2.33 V
Pin assignment DC MEAS connector
If the measured value is more than 10% above or more than 10% below the stated nominal value, the
motherboard must be replaced.