R&S ZVB
Troubleshooting
1145.1078.82 3.55
E-1
Group 1:
General Functions
Functions
Service
function
Data Serv.
Lev.
Enable/disables the arbitrary mode settings in
the port configuration (e.g. LowNoise/
LowDistortion) independently of the model
1.0.0.0.X
0
→
Use disabled
1
→
Use enabled
2
Enables/disables the peak detector and RMS
detector independently of the model
1.0.0.1.X
0
→
Use disabled
1
→
Use enabled
2
Activates or deactivates the setting of measured
values to default values when status messages
are issued due to HW faults
1.1.0.2.X
0
→
Deactivate default values
1
→
Activate default values
0
Activates or deactivates factory calibration for
the active setup
1.1.0.3.X
0
→
Deactivate factory calibration
1
→
Activate factory calibration
0
Group 2:
General Functions
Functions
Service function
Data Serv.
Lev.
Suppress error-message box
2.0.0.0.X
X=0
→
MSG box is output (default state)
X=1
→
MSG box is suppressed (error is
nevertheless entered in the log file)
1
Selftest all boards
2.0.0.5.0.0
0, 1 or 2
Network controller
Functions
Service function
Data Serv.
lev.
ND1:Read Temp (addr.:68)
2.1.2.3
1
ND2:Read Temp (addr.:68)
2.2.2.3
1
ND[1..2] Selftest
2.[1-2].2.5.0.0
0, 1 or 2
NA1:Read Temp (addr.:69)
2.1.3.3
1
NA2:Read Temp (addr.:69)
2.2.3.3
1
Summary of Contents for ZVB 4
Page 57: ...Instrument Construction and Function Description R S ZVB 1145 1078 82 I 3 2 E 1 ...
Page 133: ...1145 1078 82 5 5 E 1 Spare Parts List Mechanical Drawings ...
Page 142: ...1145 1078 82 5 15 E 1 Block Circuit Diagram ...
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 148: ......
Page 149: ......
Page 150: ......
Page 151: ......
Page 152: ......
Page 153: ...1145 1078 82 5 17 E 1 Circuit diagrams ...
Page 154: ......
Page 155: ......
Page 156: ......