Remote Control Commands
R&S
®
NRPxxT(N)
89
User Manual 1177.6023.02 ─ 08
Parameters:
<count>
Range:
1 to 65536
*RST:
4
Example:
AVER:COUN 1
Manual operation:
See
[SENSe<Sensor>:]AVERage:COUNt:AUTO
<state>
Sets the mode for determining the average count.
Parameters:
<state>
ON
Auto averaging: the averaging factor is continuously determined
and set depending on the level of power and other parameters.
OFF
Fixed filter: the previous, automatically determined averaging
factor is used.
ONCE
An averaging factor is determined by the filter automatic function
under the current measurement conditions and is then used in
the fixed filter mode.
*RST:
ON
Manual operation:
See
[SENSe<Sensor>:]AVERage:COUNt:AUTO:MTIMe
<maximum_time>
Sets an upper limit for the settling time of the auto-averaging filter if
[SENSe<Sensor>:]AVERage:COUNt:AUTO:TYPE
is set to
NSRatio
. Thus it limits
the length of the filter.
Parameters:
<maximum_time>
Range:
0.01 to 999.99
*RST:
4.00
Default unit: Seconds
Manual operation:
See
[SENSe<Sensor>:]AVERage:COUNt:AUTO:NSRatio
<nsr>
Determines the relative noise component in the measurement result for the measure-
ment modes with scalar results. These measurement modes arecontinuous average,
burst average and timeslot, provided the particular sensor supports them.
This command is only effective if the auto average calculation is enabled:
●
[SENSe<Sensor>:]AVERage:COUNt:AUTO
ON
●
[SENSe<Sensor>:]AVERage:COUNt:AUTO:TYPE
NSR
Configuring Basic Measurement Parameters