Browser-Based User Interface
R&S
®
NRPxxT(N)
49
User Manual 1177.6023.02 ─ 08
Remote command:
Zero Calibration
Performs zeroing using the signal at the sensor input. See
Note:
Turn off all test signals before zeroing. An active test signal during zeroing causes an
error.
Remote command:
CALibration<Channel>:ZERO:AUTO
Sensor Information
Displays information on the power sensor.
●
General information: manufacturer, sensor type, stock number, serial number, firm-
ware version
●
Network settings: MAC address, hostname, IP address, sensor name. A part of
them, you can configure under
Chapter 6.5.4, "System Settings"
●
Important specifications: technology, frequency range, power limits, resolution, ...
●
Calibration information
Remote command:
6.5.2
Averaging Settings
Describes the parameters for automatic averaging.
Further information:
●
Chapter 9.8.1, "Configuring Auto Averaging"
Access: main dialog of the web user interface > navigation pane > "Averaging"
............................................................................................... 50
Filter Terminal Control
Defines how the measurement results are output, also called termination control. See
also
Chapter 9.5, "Controlling the Measurement"
"Moving"
Outputs intermediate values to facilitate early detection of changes in
the measured quantity. In the settled state, that means when the
number of measurements specified by the averaging factor has been
performed, a moving average is output.
Settings