CALibration<Channel>:ZERO:AUTO
................................................................................ 110
CALibration:DATA
<caldata>
Writes a binary calibration data set in the memory of the sensor.
Parameters:
<caldata>
<block_data>
CALibration:DATA:LENGth?
Queries the length in bytes of the calibration data set currently stored in the flash mem-
ory. Programs that read out the calibration data set can use this information to deter-
mine the capacity of the buffer memory required.
Example:
CAL:DATA:LENG?
Query
57392
Response
Usage:
Query only
CALibration:TEST?
The thermal power sensors are equipped with an additional test heater.
Queries the power difference in W when measuring with the external heater enabled
and disabled.
Usage:
Query only
CALibration:TEST:DEViation?
The thermal power sensors are equipped with an additional test heater.
Queries the relative deviation of the current power difference from the reference value
stored in the calibration data set. The relative deviation is calculated as follows:
<relative
deviation>
= (<current
power
difference>
/ <reference
value>) - 1
with
<current power difference>:
Usage:
Query only
Calibrating, Zeroing