Remote Control Commands
R&S
®
NRPxxT(N)
97
User Manual 1177.6023.02 ─ 08
[SENSe<Sensor>:]SGAMma:CORRection:STATe
<state>
Enables or disables the use of the complex reflection coefficient to correct the interac-
tions of power sensor and signal source.
Parameters:
<state>
*RST:
OFF
Manual operation:
See
[SENSe<Sensor>:]SGAMma:MAGNitude
<magnitude>
Sets the magnitude of the complex reflection coefficient of the source,
Γ
source
.
Parameters:
<magnitude>
0.0
≙
ideal matched source
1.0
≙
total reflection
Range:
0.0 to 1.0
*RST:
0.0
Manual operation:
See
[SENSe<Sensor>:]SGAMma:PHASe
<phase>
Sets the phase angle of the complex reflection coefficient of the source,
Γ
source
.
Parameters:
<phase>
Range:
-360.0 to 360.0
*RST:
0.0
Manual operation:
See
9.8.3.5
I-Gamma Queries
For the current frequency, queries the complex input reflection coefficient
Γ
in
of the fol-
lowing:
●
Power sensor if the S-parameter correction is disabled.
(
[SENSe<Sensor>:]SGAMma:CORRection:STATe
OFF
)
●
S-parameter device if the S-parameter correction is enabled, S11 in
(
[SENSe<Sensor>:]SGAMma:CORRection:STATe
ON
)
[SENSe<Sensor>:]IGAMma:MAGNitude?
...........................................................................97
.................................................................................98
[SENSe<Sensor>:]IGAMma:EUNCertainty?
........................................................................98
[SENSe<Sensor>:]IGAMma:MAGNitude?
Queries the magnitude of the complex input reflection coefficient
Γ
in
.
Configuring Basic Measurement Parameters