Browser-Based User Interface
R&S
®
NRPxxT(N)
51
User Manual 1177.6023.02 ─ 08
<Slope> ← Trigger Source
Sets the polarity of the active slope of the trigger signal that is externally or internally
applied.
"Positive"
The rising edge of the trigger signal is used for triggering.
"Negative"
The falling edge of the trigger signal is used for triggering.
Remote command:
Trigger Level
Sets the trigger threshold for internal triggering derived from the test signal.
Remote command:
Trigger Delay
Sets the delay between the trigger event and the beginning of the actual measurement.
Note:
In continuous average mode, a negative trigger delay is not supported.
Remote command:
Dropout
With a positive (negative) trigger slope, the dropout time is the minimum time for which
the signal must be below (above) the power level defined by "Trigger Level". See also
Chapter 9.5.2.4, "Dropout Time"
Remote command:
Holdoff
Sets the hold-off time, a period after a trigger event during which all trigger events are
ignored. See also
Chapter 9.5.2.5, "Hold-Off Time"
Remote command:
Hysteresis
Sets the hysteresis in dB. A trigger event occurs, if the trigger level:
●
Falls below the set value on a rising slope.
●
Rises above the set value on a falling slope.
Thus, you can use this setting to eliminate the effects of noise in the signal for the edge
detector of the trigger system.
Remote command:
Settings