Preparing for Use
R&S
®
NRPxxT(N)
21
User Manual 1177.6023.02 ─ 08
Setup with a PoE injector
HOST
INTERFACE
IN:
3 V or 5 V logic
OUT: min. 2 V into 50
Ω
max. 5.3 V
TRIG2
I/0
PoE
SMART SENSOR TECHNOLOGY
NRP
1
2
3
4
5
6
7
8
Figure 3-5: Setup with a PoE injector
1
= Signal source
2
= LAN power sensor
3
= RJ-45 Ethernet connector
4, 6 = RJ-45 Ethernet cable
5
= Controlling host
7
= PoE injector
8
= AC supply
1. Connect the RF connector of the sensor to the DUT.
Chapter 3.4, "Connecting to a DUT"
2.
NOTICE!
Risk of sensor damage. Use only PoE power sourcing equipment (PSE)
according to IEEE standards 802.3af or IEEE 802.3at.
Otherwise your power sensor can get damaged.
Connect the RJ-45 Ethernet connector of the sensor to the output of the PoE injec-
tor.
3. Connect the PoE injector to a power supply.
4. Connect the controlling host to the input of the PoE injector.
5. Establish a network connection between the power sensor and the controlling host.
3.6.3.2
Establishing a Connection to the Network
There are two methods to establish a network connection:
●
Power sensor and controlling host are connected to a common network
(infrastructure network).
●
Power sensor and controlling host are connected only over the switch
(peer-to-peer network).
In both cases, you can address the LAN power sensor as follows:
●
Chapter 3.6.3.3, "Using Hostnames"
Connecting to a Controlling Host