Remote Control Commands
R&S
®
NRPxxT(N)
86
User Manual 1177.6023.02 ─ 08
100
μ
s is the time for switching the chopper phase.
Using
[SENSe<Sensor>:][POWer:][AVG:]FAST
ON
, you can accelerate the mea-
surement as follows:
●
Chopper is disabled.
●
Average count is set to 1, no matter which average count you have set.
Thus, the overall measurement time is only defined by the aperture time, and the mea-
surement time for a fast measurement is calculated as follows:
MT = APER
Remote commands:
[SENSe<Sensor>:][POWer:][AVG:]APERture
......................................................................86
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:CLEar
...............................................................86
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:COUNt?
............................................................86
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:DATA?
.............................................................. 87
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:SIZE
.................................................................87
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:STATe
...............................................................87
[SENSe<Sensor>:][POWer:][AVG:]FAST
............................................................................ 87
[SENSe<Sensor>:][POWer:][AVG:]SMOothing:STATe
.......................................................... 88
[SENSe<Sensor>:][POWer:][AVG:]APERture
<integration_time>
Sets the duration of the sampling window in continuous average mode. During this
time interval, the average signal power is measured.
Parameters:
<integration_time>
Range:
0.5e-3 to 0.3
*RST:
0.005
Default unit: Seconds
Example:
APER 0.005
Manual operation:
See
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:CLEar
Clears the contents of the result buffer continuous average mode.
Example:
BUFF:CLE
Usage:
Event
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:COUNt?
Available in continuous average mode.
Queries the number of results that are currently stored in the result buffer.
Example:
BUFF:COUN?
Usage:
Query only
Configuring the Measurement Modes