Remote Control Basics
R&S
®
NRPxxT(N)
155
User Manual 1177.6023.02 ─ 08
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
0
Sensor waiting for trigger
0
0
0
0
0
0
0
0
0
0
0
0
0
0
+
Figure 11-11: Operation trigger status register
Querying the register:
●
STATus:OPERation:TRIGger:CONDition?
●
STATus:OPERation:TRIGger[:SUMMary][:EVENt]?
Table 11-13: Used operation trigger status bits and their meaning
Bit no.
Short description
Bit is set if
1
Sensor waiting for trigger
Sensor is waiting for a trigger event. When the trigger event
occurs, the sensor changes into the measuring state.
11.3.8.4
Operation Sense Status Register
The
CONDition
register contains information whether a power sensor is being initial-
ized. The
EVENt
register contains information whether an initialization was started or
completed since the last query.
A power sensor is initialized if:
●
Supply voltage is switched on (power-up).
●
Sensor was connected.
●
Reset was performed using:
–
–
Status Reporting System