GUI Reference
R&S
®
CMW-B400B/-B405A
35
User Manual 1176.8116.02 ─ 04
Fig. 4-7: Single tone measurement settings
.........................................................................................35
...............................................................................................35
Filter > Low Pass, High Pass, Weighting, Variable Bandpass
...................................... 36
....................................................................................... 36
........................................................................................ 36
Statistic Count
Configures the statistic count for single tone measurements. This parameter can also
be configured within the general settings, see
chapter 4.3.1, "General Settings (Ana-
chapter 4.3.2, "General Settings (Digital)"
Remote command:
CONFigure:AUDio:MEAS<i>:ANALog:SINGletone:SCOunt
CONFigure:AUDio:MEAS<i>:DIGital:SINGletone:SCOunt
Filter > Distortion Frequency
The "Distortion Frequency" is the reference frequency of the single tone measurement.
Set this parameter to the fundamental frequency of the AF signal. If the measurement
is coupled to a generator, the frequency configured within the generator settings is
used and displayed here for information.
Remote command:
Filter > Window Function
Selects a window function for the single tone spectrum measurement. The spectrum
measurement uses a fast Fourier transformation. The window function is applied to the
raw data in the time domain, before performing the fast Fourier transformation. This is
done in order to reduce typical errors like spectral leakage and level errors.
The following table assists you in selecting the appropriate window function.
Audio Measurement Settings