GUI Reference
R&S
®
CMW-B400B/-B405A
36
User Manual 1176.8116.02 ─ 04
Window Type
Spectral Leakage
Amplitude Accuracy
Frequency Resolution
Rectangular
Poor
Poor
Best
Hamming
Fair
Fair
Good
Hann (Hanning)
Good
Fair
Good
Blackman-Harris
Best
Good
Fair
Flat-Top
Good
Best
Poor
Remote command:
Filter > Low Pass, High Pass, Weighting, Variable Bandpass
The settings configure the individual filter stages of the AF analyzer. For a description
of the filter types, see
chapter 3.3, "AF Filter Settings"
Remote command:
Analog measurement:
CONFigure:AUDio:MEAS<i>:ANALog:SINGletone:FILTer:LPASs
CONFigure:AUDio:MEAS<i>:ANALog:SINGletone:FILTer:HPASs
CONFigure:AUDio:MEAS<i>:ANALog:SINGletone:FILTer:WEIGhting
CONFigure:AUDio:MEAS<i>:ANALog:SINGletone:FILTer:BANDpass
Digital measurement:
CONFigure:AUDio:MEAS<i>:DIGital:SINGletone:FILTer:LPASs
CONFigure:AUDio:MEAS<i>:DIGital:SINGletone:FILTer:HPASs
CONFigure:AUDio:MEAS<i>:DIGital:SINGletone:FILTer:WEIGhting
CONFigure:AUDio:MEAS<i>:DIGital:SINGletone:FILTer:BANDpass
Limits (Analog Measurement)
This section configures upper or lower limits for most single tone results displayed in
the result table.
By default all limits are disabled.
Remote command:
CONFigure:AUDio:MEAS<i>:ANALog:SINGletone:LIMit
Limits (Digital Measurement)
This section configures upper or lower limits for most single tone results displayed in
the result table.
By default all limits are disabled.
Audio Measurement Settings