Command Reference
R&S
®
CMW-B400B/-B405A
70
User Manual 1176.8116.02 ─ 04
<BandRmsUpp>
Upper limit for bandpass level RMS results
Range:
0 FS to 1 FS
*RST:
1 FS
Default unit: FS
<BandRmsEnab>
OFF | ON
Disables or enables the upper limit for bandpass level RMS
results
*RST:
OFF
<THDPercUpp>
Upper limit for THD results
Range:
0 % to 100 %
*RST:
0.1 %
Default unit: %
<THDPercEnab>
OFF | ON
Disables or enables the upper limit for THD results
*RST:
OFF
<THDPlusNUpp>
Upper limit for THD + N results
Range:
0 % to 100 %
*RST:
0.5 %
Default unit: %
<THDPlusNEnab>
OFF | ON
Disables or enables the upper limit for THD + N results
*RST:
OFF
<SNRLow>
Lower limit for SNR results
Range:
0 dB to 100 dB
*RST:
54 dB
Default unit: dB
<SNREnab>
OFF | ON
Disables or enables the lower limit for SNR results
*RST:
OFF
Example:
See
Firmware/Software:
V3.2.10
Manual operation:
See
"Limits (Digital Measurement)"
CONFigure:AUDio:MEAS<i>:ANALog:SPECtrum:DMODe
<DiagramMode>
CONFigure:AUDio:MEAS<i>:DIGital:SPECtrum:DMODe
<DiagramMode>
Enables or disables the evaluation of the single tone spectrum results and shows or
hides the spectrum diagram.
Audio Measurements