Command Reference
R&S
®
CMW-B400B/-B405A
69
User Manual 1176.8116.02 ─ 04
<THDPercUpp>
Upper limit for THD results
Range:
0 % to 100 %
*RST:
0.1 %
Default unit: %
<THDPercEnab>
OFF | ON
Disables or enables the upper limit for THD results
*RST:
OFF
<THDPlusNUpp>
Upper limit for THD + N results
Range:
0 % to 100 %
*RST:
0.5 %
Default unit: %
<THDPlusNEnab>
OFF | ON
Disables or enables the upper limit for THD + N results
*RST:
OFF
<SNRLow>
Lower limit for SNR results
Range:
0 dB to 100 dB
*RST:
54 dB
Default unit: dB
<SNREnab>
OFF | ON
Disables or enables the lower limit for SNR results
*RST:
OFF
Example:
See
Firmware/Software:
V3.2.10
Manual operation:
See
CONFigure:AUDio:MEAS<i>:DIGital:SINGletone:LIMit
<WeightRmsUpp>,
<WeigtRmsEnab>, <BandRmsUpp>, <BandRmsEnab>, <THDPercUpp>,
<THDPercEnab>, <THDPlusNUpp>, <THDPlusNEnab>, <SNRLow>,
<SNREnab>
Configures the limits for digital single tone measurements.
Parameters:
<WeightRmsUpp>
Upper limit for weighted level RMS results
Range:
0 FS to 1 FS
*RST:
1 FS
Default unit: FS
<WeigtRmsEnab>
OFF | ON
Disables or enables the upper limit for weighted level RMS
results
*RST:
OFF
Audio Measurements