The measurement could not be started or continued because no trigger event was
detected.
●
7 (Acquisition Error)
:
The R&S
CMW could not properly decode the RF input signal.
●
8 (Sync Error)
:
The R&S
CMW could not synchronize to the RF input signal.
●
9 (Uncal)
:
Due to an inappropriate configuration of resolution bandwidth, video bandwidth or
sweep time, the measurement results are not within the specified data sheet limits.
●
15 (Reference Frequency Error)
:
The instrument has been configured to use an external reference signal but the ref-
erence oscillator could not be phase locked to the external signal (e.g. signal level
too low, frequency out of range or reference signal not available at all).
●
16 (RF Not Available)
:
The measurement could not be started because the configured RF input path was
not active. This problem may occur e.g. when a measurement is started in com-
bined signal path mode and the master application has not yet activated the input
path. The LEDs above the RF connectors indicate whether the input and output
paths are active.
●
17 (RF Level not Settled) / 18 (RF Frequency not Settled)
:
The measurement could not be started because the R&S
CMW was not yet ready
to deliver stable results after a change of the input signal power / the input signal
frequency.
●
19 (Call not Established)
:
For measurements: The measurement could not be started because no signaling
connection to the DUT was established.
For DAU IMS service: Establishing a voice over IMS call failed.
●
20 (Call Type not Usable)
:
For measurements: The measurement could not be started because the estab-
lished signaling connection had wrong properties.
For DAU IMS service: The voice over IMS settings could not be applied.
●
21 (Call Lost)
:
For measurements: The measurement was interrupted because the signaling con-
nection to the DUT was lost.
For DAU IMS service: The voice over IMS call was lost.
●
23 (Missing Option)
:
The ARB file cannot be played by the GPRF generator due to a missing option.
●
24 (Invalid RF Setting)
:
The desired RF TX level or RF RX reference level could not be applied.
●
26 (Resource Conflict)
:
The application could not be started or has been stopped due to a conflicting hard-
ware resource or software option that is allocated by another application.
Stop the application that has allocated the conflicting resources and try again.
●
27 (No Sensor Connected)
:
The GPRF External Power Sensor measurement could not be started due to miss-
ing power sensor.