Command Reference
R&S
®
CMW-B400B/-B405A
66
User Manual 1176.8116.02 ─ 04
Manual operation:
See
CONFigure:AUDio:MEAS<i>:ANALog:SINGletone:FILTer:DFRequency
<DistortionFreq>
CONFigure:AUDio:MEAS<i>:DIGital:SINGletone:FILTer:DFRequency
<DistortionFreq>
Specifies the distortion frequency for single tone measurements (fundamental fre-
quency of the AF signal).
The setting command is only relevant if the measurement is not coupled to an audio
generator. Otherwise the distortion frequency is set automatically according to the gen-
erator settings.
Parameters:
<DistortionFreq>
Range:
1 Hz to 10.5E+3 Hz
*RST:
1000 Hz
Default unit: Hz
Example:
See
Firmware/Software:
V3.2.10
Manual operation:
See
"Filter > Distortion Frequency"
CONFigure:AUDio:MEAS<i>:ANALog:SINGletone:FILTer:WFUNction
<WindowFunct>
CONFigure:AUDio:MEAS<i>:DIGital:SINGletone:FILTer:WFUNction
<WindowFunct>
Selects a window function for the single tone spectrum measurement.
Parameters:
<WindowFunct>
RECT | HAMM | HANN | BHAR | FTOP
RECT
: Rectangular window
HAMM
: Hamming window
HANN
: Hann window
BHAR
: Blackman-Harris window
FTOP
: Flat-Top window
*RST:
RECT
Example:
See
Firmware/Software:
V3.2.20
Manual operation:
See
CONFigure:AUDio:MEAS<i>:ANALog:SINGletone:FILTer:LPASs
<Frequency>
CONFigure:AUDio:MEAS<i>:DIGital:SINGletone:FILTer:LPASs
<Frequency>
Configures the cutoff frequency of the lowpass filter for single tone measurements.
Audio Measurements