CONFigure:AUDio:MEAS<i>:ANALog:SCONdition
<StopCondition>
CONFigure:AUDio:MEAS<i>:DIGital:SCONdition
<StopCondition>
Specifies whether the measurement is stopped after a failed limit check or continued.
OLFA
means that the measurement is stopped and reaches the
RDY
state as soon as
one of the results exceeds the limits.
Parameters:
<StopCondition>
NONE | OLFA
NONE:
Continue measurement irrespective of the limit check
OLFA:
Stop measurement on limit failure
*RST:
NONE
Example:
See
Firmware/Software:
V3.2.10
Manual operation:
See
CONFigure:AUDio:MEAS<i>:ANALog:SCOunt
<StatCountSing>, <StatCountMult>,
<StatCountFFTN>
CONFigure:AUDio:MEAS<i>:DIGital:SCOunt
<StatCountSing>, <StatCountMult>,
<StatCountFFTN>
Specifies the statistic counts for all audio signal types. The statistic count is equal to
the number of measurement intervals per single shot.
See also: "Statistical Settings" in the R&S
CMW user manual, chapter "Remote Con-
trol"
Parameters:
<StatCountSing>
Single tone measurements statistic count
Range:
1 to 1000
*RST:
20
<StatCountMult>
Multitone measurements statistic count
Range:
1 to 1000
*RST:
20
<StatCountFFTN>
FFT noise measurements statistic count
Range:
1 to 1000
*RST:
20
Firmware/Software:
V3.2.10
Manual operation:
See
CONFigure:AUDio:MEAS<i>:ANALog:PCOupling
<PathCoupling>
Configures the DC component handling for the analog input path.